Access this content:

Abstract

Background

Measurement of respiratory function is important both in epidemiological studies and for the diagnosis of respiratory diseases in early life. Oscillometry requires minimal subject cooperation and is promising for the assessment of respiratory function in young children.

Aim

To assess the feasibility and reproducibility of oscillometric measurements in 3-year-old children.

Methods

Measurements were performed on 242 children who participated in the 3-year visit of the SEPAGES cohort, median (Q1, Q3) age: 3.1yr (3.0, 3.2); height: 95.4cm (95.0, 97.5), using the tremoFlo device with a 7 – 41Hz waveform. Valid tests were defined as per recommendations. Resistance at 7Hz (R7) data with a CV<15% were considered reproducible. We compared the individual mean of R7 based on all technically acceptable measures (R7_all) to the one based on the 2 first measures (R7_2) and computed the intraclass correlation coefficient (ICC) between R7_all and R7_2.

Results

225 children (94.9 %) had at least 2, and 209 (88.2%) at least 3 technically acceptable R7 measurements with a CV<15%. Of the latter, 197 (81.4%) did not require any measurement exclusion. The overall CV of R7 was: 6.7% (4.1, 10.1). The median (Q1-Q3) difference between R7_all and R7_2 was -0.006 (-0.029, 0.018) cmH2O.s/L, with an ICC=0.97.

Conclusions

Our preliminary analysis demonstrates excellent feasibility and reproducibility of oscillometry in 3-year-old children. These data suggest that 2 reproducible measurements may yield a comparable outcome vs. the recommended 3.

Funding

PNREST Anses, 2018/1/264 ; ERS CRC-2013-02 INCIRCLE (International Collaboration to Improve Respiratory Health in Children)



The articles include uses of tremoflo products or describes research devices that may not have been cleared by FDA
Source

Bayat S, Boudier A, Lyon-Caen S, Quentin J, Gioria Y, Slama R, et al. Feasibility and reproducibility of oscillometry for measuring respiratory function in young children: the SEPAGES cohort. Eur Respir J [Internet]. 2020 Sep 7 [cited 2020 Nov 2];56(suppl 64)

Contact us
Interested in THORASYS products? Contact us to get a personalized quote.