Accéder au contenu:

Résumé de recherche

The forced oscillation technique (FOT) is gaining clinical acceptance, facilitated by more commercial devices and clinical data. However, the effects of variations in testing protocols used in FOT data acquisition are unknown. We describe the effect of duration of data acquisition on FOT results in subjects with asthma, chronic obstructive pulmonary disease (COPD) and healthy controls.

FOT data were acquired from 20 healthy, 22 asthmatic and 18 COPD subjects for 60 s in triplicate. The first 16, 30 and 60 s of each measurement were analysed to obtain total, inspiratory and expiratory resistance of respiratory system (Rrs) and respiratory system reactance (Xrs) at 5 and 19 Hz.

With increasing duration, there was a decrease in total and expiratory Rrs for healthy controls, total and inspiratory Rrs for asthmatic subjects and magnitude of total and inspiratory Xrs for COPD subjects at 5 Hz. These decreases were small compared to the differences between clinical groups. Measuring for 16, 30 and 60 s provided ≥3 acceptable breaths in at least 90, 95 and 100% of subjects, respectively. The coefficient of variation for total Rrs and Xrs also decreased with duration. Similar results were found for Rrs and Xrs at 19 Hz.

FOT results are statistically, but likely minimally, impacted by acquisition duration in healthy, asthmatic or COPD subjects.

Source

Watts JC, Farah CS, Seccombe LM, Handley BM, Schoeffel RE, Bertolin A, et al. Measurement duration impacts variability but not impedance measured by the forced oscillation technique in healthy, asthma and COPD subjects. ERJ Open Res. 2016;2:00094–2015.

Nous contacter
Dites-nous comment nous pouvons vous aider. Nous serions heureux d'avoir de vos nouvelles !
Vous êtes intéressés par les produits THORASYS ? Contactez-nous pour obtenir un devis personnalisé.
Formulaire soumis avec succès !
Nous vous contacterons sous peu.